Frequently Asked Question

How to interface with Tango
Last Updated 6 years ago

PhysioFlow software 2.8.0 and later is compatible with SunTech Tango+ and SunTech Tango M2. For this you need to connect the Tango to the computer running the PhysioFlow software using an RS-232 cable, and a USB/Serial converter if the computer does not have a serial port.

Tango configuration
  1. When the operating screen is displayed, press the SELECT button once. This will bring up the MAIN MENU screen.
  2. Using the UP or DOWN arrows, highlight MONITOR SETUP and press the SELECT button.
  3. Using the UP or DOWN arrows, highlight STRESS SYSTEM and press the SELECT button.
  4. Using the UP or DOWN arrows, scroll through the list until CUSTOM is highlighted and press the SELECT button to confirm the choice.
  5. Using the UP or DOWN arrows, highlight PROTOCOL and press the SELECT button.
  6. Using the UP or DOWN arrows, scroll through the list until STANDARD is highlighted and press the SELECT button to confirm the choice.
  7. Using the UP or DOWN arrows, select EXIT to return to the operating screen.
Starting a Measurement
  1. Start the Physioflow software
  2. If the tango is properly detected, it is displayed in the "Detected devices" list.
    image
  3. Click on the "Start monitoring" button on the PhysioFlow Device (Enduro/QLink)
  4. When asked if you want to use the tango for this measurement, answer "yes"
    image

Please Wait!

Please wait... it will take a second!